Sirona Dental Systems Gmbh - Sirona ORTHOPHOS SL 2D Service Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH

Service Manual ORTHOPHOS SL
64 95 258 D3632
D3632.076.01.03.02 09.2016
Marking of focal spot:
Source-skin distance
Automatic exposure
blocking:
Equipment class:
Protection class
Class I device
Degree of protection against
electric shock:
Degree of protection against
ingress of water:
Year of manufacture:
Operating mode:
Long-term power output:
Anode material:
Exposure parameters for
determining leakage
radiation:
X-ray tube
Siemens SR 90/15 FN
> 200 mm (8")
The duration of automatic exposure
blocking (cooling period) depends on the
set kV/mA level and the actual exposure
time. Depending on the tube load,
interval times of 8 s to 300 s are
automatically set by the system.
Example: For program P1 with exposure
data of 84 kV / 12 mA and a radiation
time of 14.1 s, the pause duration is 150
s.
IPX0
Type B device
Ordinary equipment (without protection
against ingress of water)
(on the rating plate)
Continuous operation
200 W
Tungsten
2mA / 90 kV
3 Unit description
3.1 Hardware
41