8 Service routines
8.1 Perform service routines via the control panel
284
S017: Test step 8
8.1.14.5
Selecting the kV/mA level series
Selection field
1
* Factory setting
1. Call service routine S017.8.
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use arrow keys (A) to select the required setting in selection field 1
(see table).
Once the required setting has been selected, the Memory key (B)
lights up.
3. Save the setting.
4. Exit the service routine.
Service Manual ORTHOPHOS SL
Code
Function
1E
16 mA series for pan
16 mA series for Ceph
2E
8 mA series for pan
16 mA series for Ceph
Sirona Dental Systems GmbH
64 95 258 D3632
D3632.076.01.03.02 09.2016