6 Troubleshooting
6.1 Error logging memory
System time
Entry type
Entry data [Message]
Entry data [Error]
140
6.1.1
Example of error logging data
2006-03-06, 20:13:02
[Message]
[Message + val]
[Error]
[DeviceError]
[Error Sidexis]
[Stringname]
[Stringsegment]
[RTC Date / Time Change]
[PC Date / Time]
[Update]
Self-test: Successful
Recording started
Value: 9000
Recording stopped
Recording canceled
Termination state
Value: 0
Value: 1
Value: 2
Image transfer started
Image transfer stopped
Logbook started
Image state switched to Released
Other entry data which document the occurrence of a rescue event include:
● Image state switched to Rescue
● Rescue request Sidexis Error
● Rescue request Sidexis TrackEpilogue
● Rescue request Sidexis Timeout
These entry data may also occur after "Recording stopped" or "Cancel" and
indicate an exceptional circumstance. You can supply important information for
error diagnosis in coordination with the Sirona Customer Service Center.
E6 07 06
ERR_DX7_TTP_LOST
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS SL
System time (clock on DX11)
General system event
General system event with additional
value
Error event
Data for error event on a module
Network error event
Free status texts
Additional data (string name)
Date and time of a SIDEXIS PC
Date and time of the DX11 set
Information on the unit software update
Self-test completed successfully
Start of a recording
Sequence ID of the recording
End of an exposure
Exposure cancellation
Reason for ending exposure
Exposure completed
Exposure cancellation by user
Exposure cancellation due to internal
error
Start of image data transfer
End of image data transfer
Corresponds to unit switch-on
Exposure has been delivered to
SIDEXIS 4 and confirmed by SIDEXIS 4.
Error code
Clear text error display