8.4 kV – verification / Exposure Time Verification
1.
A
P1
14.1 64 8
Prog.
S
T
R
7.
Preparing the measurement
Attach the Mult-O-Meter sensor in the middle of the du -
1.
al sensor.
2.
Switch the unit on via the switch A (see also Operating
Instructions)
The X-Ray radiation indicator (B) lights up briefly.
After approx. 2 seconds, the green LED (C) in the
upper part of the control panel lights up. This LED
remains lit as long as the unit is on.
The start screen appears on the touchscreen of
the Easypad and the system's self-adjustment
routine starts running (for approx. 1 minute). The
rotating element rotates briefly clockwise and
counterclockwise. The diaphragm moves into
position. The forehead and temple supports on the
panoramic unit open and close and then stop
moving in fully opened position.
On completion of the self-adjustment routine, the
main menu appears on the touchscreen. Help
message H301 prompts you to move the unit into
the starting position.
3.
Press the R key.
The unit moves to its starting position.
60 51 416 D 3352
D 3352.031.03.18.02
4.
B
C
kV
mA
8.4 kV – verification / Exposure Time Verification
A
-
60kV/8mA
-
0.5s
4.
Call the Service menu and the Service routine S002.3
(see Service Manual).
5.
Use the arrow keys (A) in selection field 1 to select the
kV/mA level 62kV/8mA.
6.
Use the arrow keys (A) in selection field 2 to select the
radiation time 0.5 s.
Performing measurements
7.
Initiate the radiation. Hold the release button pressed
until the set radiation time has expired.
CAUTION! Activating the release button triggers
X-rays
8 Startup for USA/Canada only