7 Adjusting/calibrating the unit
7.1 General information about unit adjustment and calibration
162
7.1.3
Enabling exposure readiness
Enabling exposure readiness
To take an exposure in SIDEXIS XG, the system must first be made ready
for exposure.
✔
"Service functions"
GALILEOS exposure readiness
Call the
✔
Select the corresponding tab.
➢ Click on the
The exposure window opens in SIDEXIS XG. It indicates the
current status of exposure readiness.
The service routine used for the corresponding exposure is
displayed on the control panel, along with the specific exposure
parameters.
7.1.4
Taking an exposure
✔
Taking an exposure
Call the
"Service functions"
Taking an exposure in GALILEOS
✔
Select the corresponding tab.
✔
SIDEXIS XG must be ready for exposure.
1. Press the R key to move the unit into the starting position.
The procedure is complete as soon as the
status message is displayed on the touchscreen (Easypad only)
and the service routine exposure parameters are visible.
2. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
7.1.5
Save values
Save values
✔
Adjustment or calibration must be OK, i.e. the values in the menu's
text boxes must be equal to zero.
➢ To save the adjustment or calibration values, click the
button.
The adjustment or calibration is saved.
The saved adjustment or calibration is identified in the structure
tree by a check mark (2D adjustment) or a green traffic light
symbol with a check mark (3D adjustment).
menu.
"Image acquisition"
button.
menu.
Sirona Dental Systems GmbH
Service Manual GALILEOS
"Ready for exposure"
"Save values"
61 25 665 D3437
D3437.076.01.15.02 06.2012