Preparing the exposure

7.4
Pushing on the camera support
➢ Push the camera support onto the camera as illustrated.
Pushing on the camera support
Positioning the camera
CAUTION
Hot surface!
The prism of the camera is preheated in the camera holder. The
surface temperature may be as high as 50°C. This may cause an
unpleasant heat sensation on contact with a person's skin or mucous
membrane. These temperatures will not damage the skin or mucosal
membrane.
1. Position the camera over the teeth to be scanned.
Supporting the 3D camera
2. Support the camera with the front part of the camera support on a
tooth so that you can hold it quietly during the acquisition phase.
NOTE
Powder on the surface of the prism
If the prism touches powdered surfaces, then powder usually remains
on the prism surface and generates dark spots in the image.
The powder can be wiped off from the prism with a soft cloth.
7 Operation
39