9 Adjusting / calibrating the system

9.5 Sensor calibration

1.
2.
5.
6.
9.5
Sensor calibration
1. Go to the
sub-menu.
SENSOR
2. Make SIDEXIS XG ready for exposure:
Click
IMAGE ACQUISITION
The exposure dialog box showing the exposure status
appears in SIDEXIS XG.
Service routine S010.14 is displayed on the control
panel.
3. Press the R key to move the unit back to the starting
position.
4. Press the release button. Hold down the release button
until image acquisition is completed and the acoustic
signal for end of exposure (double beep) sounds − this
takes approx. 2 − 3 minutes.
5. The acquired images are transferred from the unit to
the Reconstruction and Control Unit (RCU).
This may take 2 − 3 minutes.
The evaluation of the sensor calibration is then dis-
played in the message window.
28
Maintenance Instructions GALILEOS
3.
UNTIL UNIT IS IN STARTING POSITION
4.
i
NOTE
If the information in the message window indicates that cal-
ibration is not ok and/or not possible, keep repeating the
procedure starting with Step 2 until calibration is ok and/or
possible.
If you have repeated the procedure three times and still
have not attained a positive result, please contact the
SIRONA Customer Service Center.
6.
If calibration is ok and/or possible (see above), click
to save the calibration.
SAVE VALUES
7.
Sensor calibration is now completed.
Continue the calibration procedure with iris calibration.