Displays And Help Messages During Adjustment/Calibration - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
Help messages
Status messages
63 03 510 D3352
D3352.076.03.11.02 09.2016
7.1.1
Displays and help messages during adjustment/
calibration
The most frequent help and status messages during calibration are listed
below.
H3 01: Move unit to starting position, press the R key.
H3 07: Remove the occlusal bite block.
H3 07 XG 3D / 3D ready
H3 21: Close the door.
H4 02: Plug the sensor into the ceph slot (sensor missing).
H401/402 XG 3D
H4 03:SIDEXIS XG is not ready for exposure, make unit ready for
exposure.
H4 04:Plug the sensor into the ceph slot (incorrect sensor plugged in).
H406/404 XG 3D
H4 06: Move cephalometer to starting position, press the R key.
H5 01: Turn cephalometer into lateral position.
H501/502 XG 3D
H5 02: Turn cephalometer into AP or PA position.
H6 10: Wait until preview image is acknowledged.
H610 XG 3D
● Ready for exposure
XG 3D status messages
● Exposure not possible
● Please wait
● Ready for exposure in XX seconds
● Exposure is performed
If error message E1 11 20 is displayed on the control panel and/or the
remote control during the calibration process, this does not necessarily
indicate an equipment error. This error message only indicates that the
adjustment or calibration data of the unit is incomplete at this point.
Acknowledge the error message with the R key, if applicable, and
continue the adjustment or calibration procedure.
For assistance with other help messages or error messages displayed
during the adjustment or calibration process, please refer to the section
of these instructions entitled Messages [ → 88].
7 Adjusting and calibrating the unit
7.1 General information about unit adjustment and calibration
187