Malfunctions

7.1

Help messages

When working with the unit, auxiliary messages are displayed for certain
actions (e.g. press H301 for the return key), which call for the user to
perform a specific action. These auxiliary messages are listed below. If
an error occurs, error messages are output starting with "E" followed by
5 digits, see "Error description" [ → 63].
The unit is switched on and ready for operation.
1. Press the release button.
The message H3/H4 xx appears.
2. See list below about how to proceed to make the system ready for
exposure.
H301 – R button, move into starting position
H3 01 – R key, move into starting position
The rotating unit is not in the starting position.
Moving into the starting position
CAUTION
Pressing the R key moves the unit to the starting position.
A patient positioned in the unit may be injured by moving parts.
➢ Check that a patient is not positioned in the unit before moving it to
the starting position.
➢ Press the R key.
The unit travels to the starting position.
H320 – R button, confirm exposure data
H3 20 – R key, confirm exposure data
The exposure data have not been acknowledged yet.
➢ Press the R key.
Exposure data are confirmed.
H321 – Close the door
H3 21 – Close door
Check door contact of the X-ray room.
➢ Close the door to the X-ray room.
The contact switch on the door is closed.
H401 – Plug sensor into PAN slot
H4 01 – Inserting a sensor in the PAN slot
The sensor is not in the appropriate slot for the selected exposure type.
➢ Plug the sensor into the PAN sensor slot.
The program sequence is continued.
H403 – Switch SIDEXIS to ready for exposure state
H4 03 – Switching SIDEXIS to ready for exposure state
SIDEXIS is not ready for exposure.
➢ Switch SIDEXIS to a ready for exposure state, see the SIDEXIS user
manual.