Power Meter Operation

Setting Offsets

Setting Offsets

Setting Channel Offsets

The power meter can be configured to compensate for a signal loss or gain in your test
setup (for example, to compensate for the loss of a 10 dB attenuator). This gain or
loss, referred to as a channel offset, is added to the measured power before the result
is displayed.
Offsets are entered in dB. The allowable range of values is -100 dB to +100 dB. A
positive value compensates for a loss, and a negative value compensates for a gain.
If either a channel or a display offset is set then "Ofs" is displayed.
To enter a channel offset which compensates for a loss or gain:
System
1.
Press
Inputs
2.
Press
Input Settings
3.
Press
Offset
window. Modify this offset (see below) as desired.
Use
n
positioned.
Use
n
4.
To confirm your choice press
5.
Offset Off On
using
Offset
Channel offsets can be disabled and re-enabled simply by pressing
.
Offset Off On

Setting Display Offsets

The display offset function provides a method for entering display offset values.
If either a channel or a display offset is set then "Ofs" is displayed.
To enter a display offset on the currently selected window:
Rel
1.
Press
,
Offset
2.
The power meter displays the offset value in a pop up window. Modify this
offset (see below) as desired.
2-36
.
.
. The power meter displays the offset value in a pop up
or
to modify the digit on which the cursor is currently
or
to move to other digits.
.
dB
is automatically set to "On" when a value is entered
.
.
Offset
Agilent E4418B User's Guide