3
Using P-Series Power Sensor

Setting the Trace Display

The P- Series power meter is optimized to operate in the trace display
mode when measuring a pulsed signal.
The swiftest method to get the power meter into a trace display is
described in the following procedure:
The procedure assumes you are starting from a default preset.
N O T E
1 Connect the power sensor to the power meter and connect it to the
2 Set the Channel Frequency
3 Set the Trigger to Continuous
4 Set the Display to Trace
5 Expand window and use the Gate control and Trace control menus to set
You can return to other menus later to add more accuracy on the measurement, for
N O T E
example, Channel Setup, to set the measurement average, the video average and the
bandwidth of a modulating signal.
6 When you are satisfied with the configuration you may want to save it
118
power source
the gate markers, trigger delay, and scale.
for future use. See
page 104 for further information.
"Saving and Recalling Power Meter States"