Making Measurements on the ENA
Measuring Electro-Optical Devices on the ENA
Calibrating the E/O measurement
44
Before you make a measurement, always perform an
electrical calibration of the network analyzer as described in
"Electrical calibration" on page 43.
If you make a measurement without this calibration, you will
get an error message.
Before you make your first electro- optical measurement, you
must calibrate the E/O measurement.
If you have already calibrated the E/O measurement, you can
connect your device under test and make the measurement
using the controls on the network analyzer.
Please refer to the user guide and online help for the
network analyzer for further information.
1 If you have been measuring opto- electrical or
optical/optical devices, you must first remove the
correction matrix for the active measurement type (the
active measurement type is shown at the top right of the
trace). You do this by calibrating for an E/E measurement,
as described in "Returning to Electrical Measurements on
the ENA" on page 53).
2 Press the "Macro Run" button on the front panel of the
network analzer.
Select [E/O Measurement]
The following dialog box will be displayed:
Agilent N4373A Lightwave Component Analyzer, First Edition