Calibration Procedure; Load Calibration Constants; Set Up The Analyzer - Agilent Technologies 8510XF Operating And Service Manual

Network analyzer systems 2 to 85 ghz and 2 to 110 ghz
Hide thumbs Also See for 8510XF:
Table of Contents

Advertisement

NOTE
Load Calibration
Constants

Set Up the Analyzer

Calibration Procedure

The following procedure assumes that the 1.0 mm calibration kit (85059A)
will be used to calibrate the 8510XF.
Before you can calibrate the system, you must load the appropriate
calibration constants into the system.
1. Insert the calibration data disk (included with the 1.0 mm calibration kit)
into the 8510C disk drive.
2. Press
.
[DISC] {LOAD}
The analyzer displays
3. Select
or
{CAL KIT 1}
{CAL KIT 2}
The analyzer displays
4. Select
from the display menu.
CK_1MMA1A
(This file is the recommended choice for normal operation; the other file,
, is discussed under
CK_1MMA1B
5. Select
. This loads the calibration constants into the memory
{LOAD FILE}
of the network analyzer. A screen message indicates when the load
process is complete (
6. Remove the disk from the drive.
1. Set the desired frequency range for the calibration, using the
keys.
2. Select step mode, using
3. Set the desired number of points, using
.
{NUMBER OF POINTS}
4. Set the desired averaging value, using
.
{AVERAGING ON/restart}
5. Be sure that the system impedance is set to 50
[CAL] {MORE} {SET Z0}
(If the display does not read
6. Press
or
[CAL] {CAL 1}
{CAL 2}
SELECT DATA TYPE TO LOAD
.
USE KNOB OR STEP KEYS TO SELECT A FILE
"Broadband Standards" on page
FILE 'CK_1MMA1A' LOADED FROM DISK
.
STIMULUS [MENU] {STEP}
STIMULUS [MENU]
RESPONSE [MENU]
press:
[5] [0] [X1]
50.0Ω,
8510XF Network Analyzer Systems 3-9
Operation
Calibration Procedure
.
.
3-43).
).
STIMULUS
, by pressing:
.)

Advertisement

Table of Contents
loading

This manual is also suitable for:

E7340aE7342aE7350aE7352a

Table of Contents