Operational Procedure (When Using Mechanical Calibration Kit) - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Figure 4-75
Status indication for the traces in which scalar-mixer is valid
NOTE
In scalar-mixer calibration, the normal calibration coefficient is invalid while the
frequency-offset sweep is in progress; in this case the scalar-mixer calibration coefficient is
used instead. Turning the frequency-offset sweep (
to the normal calibration coefficient; however, the information on the scalar-mixer
calibration coefficient is retained.

Operational Procedure (when using mechanical calibration kit)

NOTE
Before starting scalar-mixer calibration, verify that the frequency-offset (
Offset
Step 1. Press
Step 2. Press
Step 3. Press
Step 4. Press
Step 5. Press
Step 6. Select the test port from the list below for starting scalar-mixer calibration. In the following
steps, x and y signify the selected test ports.
Softkey
Port 1
Port 2
Port 3
Port 4
2-1 (fwd)
1-2 (rev)
1, 2 (both)
3-1 (fwd)
1-3 (rev)
1, 3 (both)
4-1 (fwd)
Chapter 4
) is
. If the frequency-offset is not valid, you cannot start scalar-mixer calibration.
ON
or
.
Mixer/Converter Calibration
.
Scalar Cal (Manual)
.
Select Ports
Function
Selects test port 1
Selects test port 2
Selects test port 3
Selects test port 4
Selects forward direction for test ports 1 and 2
Selects reverse direction for test ports 1 and 2
Selects both directions for test ports 1 and 2
Selects forward direction for test ports 1 and 3
Selects reverse direction for test ports 1 and 3
Selects both directions for test ports 1 and 3
Selects forward direction for test ports 1 and 4
Scalar-Mixer Calibration
Frequency Offset
to select the channel you want to calibrate.
.
Calibration
)
switches over
OFF
Frequency
235