Checking Calibration Kit Definitions

After an upgrade, an 8510 contains calibration kit definitions that may be significantly
different from those that were in the pre-upgraded analyzer. Ensure that the calibration
kit definitions contained in the 8510 match the calibration kit used by reloading the
definitions from disk.

Viewing Calibration Kit Labels

1. On the 85102, press
2. Check the calibration kits listed.
3. Load the calibration kit definition for the calibration kit used from disk.

Loading a Calibration Kit Definition

1. Insert the calibration kit data disk into the 8510 disk drive.
The master calibration constants disk included in the upgrade package (8510
Calibration Data disk p/n 08510-10034) contains calibration constants for many
standard calibration kits. The calibration kits are listed by model number on the disk
label. This disk can be used to load calibration kit definitions in case the calibration kit
does not have a disk.
2. Press
AUXILIAIRY MENUS TAPE (or DISC)
3. Select
[CAL KIT 1-2]
4. Select the data file to load, normally FILE 1.
5. Select the calibration kit you want to replace with the definition from disk.
a. Using the RPG knob or up/down arrow keys, select the appropriate kit.
b. Select
[LOAD FILE]
c. Press
to verify that the calibration kit loaded properly.
[CAL]
NOTE
For best measurement results, the 8510 must have the proper
calibration kit definition. If you have any doubt, reload the definition
from disk.
85103E Performance Upgrade Installation Manual
.
CAL
.
.
85103E Upgrade Package (8510A to 8510C)
Checking Calibration Kit Definitions
and select
.
[LOAD]