Error Terms

Measurement Calibration Procedure

Measurement Calibration Procedure
1. Refer to
"Measurement Calibration" on page 2-22
calibration on ES models or the enhanced response cal on ET models with the following
modifications:
2. Connect a testport cable with a female connector to PORT 2 (TRANSMISSION port on
ET models). Use the female standards (open, short, load) on PORT 1 (REFLECTION
Port on ET models) and the male standards (open, short, load) on the cable.
3. For the isolation measurement, select from the following two options:
• If you will be measuring devices with a dynamic range less than 90 dB, press:
OMIT ISOLATION
• If you will be measuring devices with a dynamic range greater than 90 dB, follow
these steps:
a. Leave the cables connected and connect impedance-matched loads to the test
ports (or reference test ports).
NOTE
If you will be measuring highly reflective devices, such as filters, use the test
device, connected to the reference plane and terminated with a load, for the
isolation standard on PORT 1 (REFLECTION port on ET models). Connect a
load to PORT 2 (TRANSMISSION port on ET models).
b. Press
Avg
c. Press
Avg
d. Press the following:
RESUME CAL SEQUENCE
Cal
ES models:
ET models:
e. Return the averaging and bandwidth to the original state of the measurement,
and press the following:
Cal
RESUME CAL SEQUENCE
ES models:
ET models:
11-4
ISOLATION DONE
AVERAGING ON
AVERAGING FACTOR
to change the IF bandwidth to 10 Hz.
IF BW
10
x1
DO BOTH FWD + REV
FWD ISOL'N
DONE | 2-PORT CAL
DONE FWD ISOL' ENH RESP
and perform the full 2-port
ISOLATION
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
.
8
x1
Service Guide