Optimizing Measurement Results
Increasing Sweep Speed
If you must view both measurement channels simultaneously (with dual channel), use the chop sweep
mode, explained next.
3. If you want to view channel 3 (or channel 4), press
dual trace display of channel 1 and channel 3 (or channel 2 and channel 4).
To return to a single-trace display, press
AUX CHAN on OFF

To Activate Chop Sweep Mode

You can use the chop sweep mode to make two measurements at the same time. For example, the analyzer
can measure S11 and S21 simultaneously. You can activate the chop mode by pressing
pressing
Cal
MORE
While Chop mode is the fastest way to measure devices, some components, such as filters with very high
attenuation, may require measurement in Alternate mode. See

To Use External Calibration

Off-loading the error correction process to an external PC increases throughput on the network analyzer.
This can be accomplished with remote-only commands. Refer to the programmer's guide for information on
how to use external calibration.

To Use Fast 2-Port Calibration (ES Analyzers Only)

With a 2-port calibration on, faster trace updates are possible by not measuring the reverse path for every
forward sweep. This is controlled by the test set switch command. This is convenient for tuning applications
because it gives a faster trace update.
When making measurements using full two-port error-correction, the following types of test set switching
can be defined by the user:
• Hold: In this mode the analyzer does not switch between the test ports on every sweep. The
measurement stays on the active port after an initial cycling between the ports. The fastest
measurements can be made by using this type of test set switching. Pressing the
to a different S-parameter measurement, or any other action which restarts a sweep will cause the test
set to switch and cycle between the ports.
5-12
Display
.
.
CHOP A and B
(or
). This will always result in a
Chan 3
Chan 4
DUAL | QUAD SETUP
"Increasing Dynamic Range" on page
or by
Preset
5-14.
key, changing
Meas