To View A Single Measurement Channel; To Activate Chop Sweep Mode; To Use External Calibration; To Use Fast 2-Port Calibration (Es Analyzers Only) - Agilent Technologies 8753ET User Manual

Network analyzers
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

Optimizing Measurement Results
Increasing Sweep Speed

To View a Single Measurement Channel

Viewing a single channel will increase the measurement speed if the analyzer's channels
are in alternate, or uncoupled mode.
1. Press
Display
DUAL | QUAD SETUP
AUX CHAN on OFF
2. Press
and
Chan 1
If you must view both measurement channels simultaneously (with dual channel), use
the chop sweep mode, explained next.
3. If you want to view channel 3 (or channel 4), press
always result in a dual trace display of channel 1 and channel 3 (or channel 2 and
channel 4).
To return to a single-trace display, press
AUX CHAN on OFF

To Activate Chop Sweep Mode

You can use the chop sweep mode to make two measurements at the same time. For
example, the analyzer can measure S11 and S21 simultaneously. You can activate the chop
mode by pressing
Preset
While Chop mode is the fastest way to measure devices, some components, such as filters
with very high attenuation, may require measurement in Alternate mode. See
Dynamic Range" on page

To Use External Calibration

Off-loading the error correction process to an external PC increases throughput on the
network analyzer. This can be accomplished with remote-only commands. Refer to the
programmer's guide for information on how to use external calibration.

To Use Fast 2-Port Calibration (ES Analyzers Only)

With a 2-port calibration on, faster trace updates are possible by not measuring the
reverse path for every forward sweep. This is controlled by the test set switch command.
This is convenient for tuning applications because it gives a faster trace update.
When making measurements using full two-port error-correction, the following types of
test set switching can be defined by the user:
• Hold: In this mode the analyzer does not switch between the test ports on every sweep.
The measurement stays on the active port after an initial cycling between the ports.
The fastest measurements can be made by using this type of test set switching.
Pressing the
Meas
other action which restarts a sweep will cause the test set to switch and cycle between
the ports.
5-12
.
to alternately view the two measurement channels.
Chan 2
.
or by pressing
Cal
5-14.
key, changing to a different S-parameter measurement, or any
DUAL CHAN on OFF
(or
Chan 3
Display
DUAL | QUAD SETUP
MORE
CHOP A and B
). This will
Chan 4
.
"Increasing