Default Cal Coefficient Adjustment
The purpose of this procedure is to keep the default of the coefficient of calibration.
Required Equipment
Calibration Kit
Procedure
Step 1. Run the adjustment program.
Step 2. Choose the Default Cal Adjustment in the Spot Adjustment.
It is possible to perform this item with the All Adjustment too.
Step 3. Connect the instrument as shown in Figure 3-15.
Figure 3-15
Default Cal Coefficient Adjustment Setup
Step 4. Follow the adjustment program instruction to update the correction constant.
Chapter 3
Description
16195B(furnished with E4991A)
Default Cal Coefficient Adjustment
Recommended Model
Adjustment
141