Performing Adjustments And Performance Tests; Setting Up The System - Agilent Technologies E4915A Service Manual

Crystal impedance / lcr meter
Table of Contents

Advertisement

Performing Adjustments and Performance Tests

The adjustments and performance tests are as follows:
1. Pre-check and Source Linearity Calibration (PRE CHECK)
2. Oscillator Calibration (OSC CAL)
3. Source Level Accuracy and Source Power Linearity Test (OSC TEST)
4. Frequency Accuracy Test (FREQ ACCY)
5. Through Calibration (THRU CAL)
6. Receiver Test (RECEIVER)
7. Isolation Test (ISOLATION)
8. Open Calibration with PI Test Fixture (OPEN CAL)
9. Short Calibration with PI Test Fixture (SHORT CAL)
10. Load Calibration with PI Test Fixture (LOAD CAL)
11. Load Test with PI Test Fixture (LOAD TEST)
12. Open Calibration with Impedance Probe (OPEN CAL 001)
13. Short Calibration with Impedance Probe (SHORT CAL 001)
14. Load Calibration with Impedance Probe (LOAD CAL 001)
15. Load Test with Impedance Probe (LOAD TEST 001)
1 Performance Tests. 2 E4916A with Option 001 only.

Setting Up the System

Performing adjustments and performance tests requires the system described in this section.
The hardware setup is shown in Figure 2-1.
2-4 Adjustments and Performance Tests