System Verification and Performance Tests
System Verification

Measurement Calibration

1. Continue with cables connected to PORT 1 and PORT 2 (REFLECTION port and
TRANSMISSION port on ET models) (see
Figure 2-9 Connections for Measurement Calibration Standards
2. Press the following:
ES models:
Cal
ET models:
Cal
TRAN/REFL ENH. RESP. REFLECTION
Port 1 Calibration
3. Connect the open that is supplied in the calibration kit to cable 1.
4. Press
FORWARD:OPEN
5. When the analyzer finishes measuring the standard, connect the short that is supplied
in the calibration kit to cable 1.
6. Press
FORWARD:SHORT
7. When the analyzer finishes measuring the standard, connect the 50Ω termination that
is supplied in the calibration kit to cable 1.
8. Press
FORWARD:LOADS
2-22
CALIBRATE MENU FULL 2-PORT
CALIBRATE MENU ENHANCED RESPONSE
.
.
.
Figure
2-9).
REFLECTION
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
Service Guide