Setting Measurement Parameters For The Power Meter Calibration - Agilent Technologies 8719ES User Manual

Network analyzer
Hide thumbs Also See for 8719ES:
Table of Contents

Advertisement

Making Mixer Measurements (Option 089 Only)
Conversion Loss Using the Frequency Offset Mode

Setting Measurement Parameters for the Power Meter Calibration

1. Connect the measurement equipment as shown in Step 1 of
Figure 2-11 Connections for R Channel and Source Calibration
CAUTION
Note that the front panel jumper between R In and R Out must remain
installed during the procedure steps that use the connections in Step 1 of
Figure
2-11.
2. From the front panel of the analyzer, set the desired receiver (RF) frequency and source
output power by pressing:
System
INSTRUMENT MODE
Start
650
Note that these are the example RF start and stop frequencies. Enter the
RF start and stop frequencies for your measurement instead.
If the LO frequency is not set to 0 Hz, press:
LO FREQUENCY
3. To select the measurement trace, press:
Meas
INPUT PORTS
The measurement trace is shown on the display.
4. Select the analyzer as the system controller:
SYSTEM CONTROLLER
Local
2-14
FREQ OFFS MENU
M/µ
Stop
950
0
x1
R
Figure
M/µ
2-11.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8719et8720es8720et8722es8722et

Table of Contents