Specications
Measurement Parameters
R (dc resistance), I (dc current)
Note
The 4349B does not contain a test voltage source. An external voltage
source for resistance measurements must be supplied by the user. The 4349B
calculates the resistance using current (I) measurement data and the test
voltage data entered into memory by the user.
Measurement Conditions
DC Test Voltage Range
0 to 5000 V in 0.1 V steps (data entry only)
Number of Test Channels
4 channels (Option 001: 2 channels)
On being triggered the 4349B simultaneously measures each channel.
Input Resistance
6
1 k
5%
Input Oset Voltage Accuracy
6
2.5 mV
Ranging
Auto and Hold (manual)
Trigger
Mode
Internal, Manual, External, and Bus
Delay Time
0 to 9.999 s in 0.001 s steps
Test Cable Length
2 meters maximum
Measurement Time Mode
10 ms, 30 ms, 100 ms, and 400 ms
Averaging
1 to 256
8-2
General Information
4349B