Limit Line Offset - Agilent Technologies ENA Series User Manual

Rf network analyzers
Hide thumbs Also See for ENA Series:
Table of Contents

Advertisement

Limit line offset

By adding a certain offset to the limit value, you can adjust the limit line so that it conforms
to the device output.
Step 1. Press
limit test function.
Step 2. Press
function.
Step 3. Press
Step 4. Press
Step 5. Press
functions correspond to each softkey.
Softkey
Stimulus Offset
Amplitude Offset
Marker Amp. Ofs.
Return
Figure 11-5
Stimulus offset
Chapter 11
or
or
to activate the trace to which you want to apply the limit test
to display the Analysis menu.
to display the softkeys for the limit test.
Limit Test
to display the limit line offset function menu. The following
Limit Line Offsets
Function
Adds a certain offset to the stimulus value of the entire segment in the
limit table. (Stimulus offset)
Adds a certain offset to the response value of the entire segment in the
limit table. (Amplitude offset)
Adds the amplitude offset by the same amount as the retrieved value of the
active marker. You can confirm the current value set for the amplitude
offset by pressing
Ends the limit line offset menu.
to activate the channel to which you want to apply the
. (Marker amplitude offset)
Amplitude Offset
Limit Test
Limit Test
381