Viewing The Wide Band Response To A Comb Signal (Inner Loop Test) - Agilent Technologies PSA Series Service Manual

Spectrum analyzer
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Hardware Options
Verifying Option 122 or 140, Wide Bandwidth Digitizer
Viewing the wide band response to a comb signal
(inner loop test):
The wideband IF is only accessible in the Basic instrument mode.
1. Press
2. Press
3. Press
4. Press
5. Press
6. Press
You will see a display of the internal comb signal that is generated on
the A7 Wideband Analog IF assembly and sent through both the Analog
IF assembly and the A32 Wide Band Digital IF assemblies. We will call
this the inner loop test since it is contained inside the wideband Analog
IF and Digital IF assemblies only. Disregarding the center tone, all
comb teeth should be within 15 dB of each other, and at least 20 dB
above the noise floor. See
180.
Figure 8-5
Comb Teeth
178
,
Mode
Basic
,
Meas Setup
IF Path
,
Input/Output
Input Port
.
IF Align Signal
and select
Signal Type
entire 80 MHz IF bandwidth (for Option 122) or 40 MHz IF
bandwidth (for Option 140).
,
Meas Setup
Res BW
,
. This selects the wideband IF path.
Wide
,
.
IF Align
The comb signal stimulates the
Comb
and type in
150kHz
Figure 8-5
below. Also see the chart on
page
Chapter 8