Adjustments and Correction Constants

Sampler Calibration Correction Constants (Test 51)

Sampler Calibration Correction Constants (Test 51)
Required Equipment and Tools
Equipment Description
1
Power Meter
Power Sensor
Adapter 3.5-mm (f) to 3.5-mm (f)
Adapter 2.4-mm (f) to 2.4-mm (f)
HP/Agilent RF Cable Set
RF Open
1. The EPM-441A and EPM-442A, with versions prior to A 2.02.00 firmware, cannot be
controlled by the analyzer. These versions do not have the 438A command set
firmware. These models may be upgraded at an Agilent Technologies service center.
Analyzer and power meter warmup time: 30 minutes.
IMPORTANT
Perform
this test.
1. Make sure the A7 Switch is in the alter position (ALT).
2. Press
PRESET FACTORY
Preset
3. Connect the equipment as in
3-58
436A, 437B, 438A, E4418A (EPM-441A) or
E4419A (EPM-442A)
8485A
part number
85052-60012
(Part of 85052B)
N/A
85131F
part number
85052-60009
(Part of 85052B)
"Raw Offset Correction Constants," on page 3-48
Preset
Figure
3-16.
8719ET/ES
8720ET/ES
8487A
N/A
part number
85056-60006
(Part of 85056A)
85133F
part number
85056-60023
(Part of 85056A)
.
8719ET/20ET/22ET
8719ES/20ES/22ES
RF Network Analyzers
8722ET/ES
before performing
Service Guide