OPEN/SHORT Response Calibration (reflection test)

In OPEN or SHORT response calibration, calibration data are measured by connecting an
OPEN or SHORT standard, respectively, to the desired test port. For frequency response,
these calibrations effectively eliminate the reflection tracking error from the test setup in a
reflection test using that port (Figure 4-3). It is also possible to carry out isolation
calibration with a LOAD standard during OPEN/SHORT response calibration. An
isolation calibration will eliminate the directivity error from the test setup in a reflection
test using that port (Figure 4-4).
Figure 4-3
1-Port error model (OPEN/SHORT response)
Figure 4-4
1-Port error model (OPEN/SHORT response + isolation)

Procedure

Step 1. Press
calibration.
Step 2. Press
Step 3. Press
Chapter 4
OPEN/SHORT Response Calibration (reflection test)
or
.
.
Calibrate
to select the channel for which you want to perform the
Calibration
107