Dosimetry

Dosimetry
A dosimeter for pulsed radiation (e.g. Mult-O-Meter 512L) is required for
dosimetry.
"Service functions"
The
The distortion phantom is clipped onto the cover of the X-ray detector
for protection against scratching.
"Dose"
1. Click the
The corresponding tab card is selected.
2. Attach the Mult-O-Meter sensor approximately in the middle of the
distortion phantom mounted on the X-ray detector.
3. Click on the
SIDEXIS XG makes the unit ready for exposure [ → 162].
Service routine S011.8 is displayed on the control panel.
4. Press the R key to move the unit back to the starting position.
5. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
7.3 Adjustment and calibration via the "Service Functions" menu
menu is called [ → 161].
tab.
"Image acquisition"
button.
7 Adjusting/calibrating the unit