13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu

Dosimetry

13.2.2.6
A dosimeter for pulsed radiation (e.g. Mult-O-Meter 512L) is required for
dosimetry.
1. In the structure tree, under
"Dose measurement"
2. Attach the Mult-O-Meter sensor in the middle of the combisensor.
3. Make SIDEXIS XG ready for exposure. [ → 206]
4. Take an exposure (85 kV/35 mA s) [ → 206].
5. Read the dose shown on the Mult-O-Meter.
6. Note down the dose for later recording.
7. If the measured dose is between 1.2 and 2.3 mGy, check the box in
the action area next to the text
.
 The traffic light symbol in front of the
element in the structure tree changes color to green (with check
mark).
8. Remove the Mult-O-Meter from the dual sensor.
13 Adjusting and calibrating the unit
"3D Adjustment/calibration"
element.
"Dose is within range of 1,2 - 2,3 mGy"
"Dose measurement"
, click on the
255