Performing Measurement Calibration; Using Flowgraphs To Identify Error Terms - Agilent Technologies E8362B Service Manual

Pna series microwave network analyzers
Hide thumbs Also See for E8362B:
Table of Contents

Advertisement

General Purpose Maintenance Procedures
Error Terms

Performing Measurement Calibration

A calibration must be performed to allow the analyzer to calculate the error terms before
they can be used as a tool:
Perform the following procedure only at a static-safe workstation, and wear a
CAUTION
grounded wrist strap.
This is important. If not properly protected against, electrostatic discharge
can seriously damage your analyzer, resulting in costly repair.
To reduce the chance of electrostatic discharge, follow all of the
recommendations outlined in
page
1-6, when performing the following calibration.
1. Connect a test cable to Port 2.
2. Perform a full 2-port calibration,
analyzer if necessary.

Using Flowgraphs to Identify Error Terms

Flowgraphs are a graphical representation of signal flow through the measurement path.
The flowgraphs in
Figure
terms associated with measurement calibration for 1-port, 2-port, 3-port, and 4-port
configurations respectively.
Figure 8-1 Flowgraph of One-Port Error Terms for Port 1
where:
E = Error term
Subscript:
D = Directivity
S = Source Match
R = Reflection Tracking
The error terms are the same for a one port measurement on Port 2 (S
8-4
"Electrostatic Discharge Protection" on
FULL SOLT 2-Port
8-1,
Figure
8-2,
Figure
PNA Series Microwave Network Analyzers
. Refer to embedded help in the
8-3, and
Figure 8-4
illustrate the error
Service Guide E8364-90026
E8362B, E8363B, E8364B
).
22

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8363bE8364b

Table of Contents