CONFigure [1] |2|3|4 Commands

The CONFigure commands are used on the specified window/measurement
to set:
– The expected power level being measured.
– The resolution of the window/measurement.
– The channel(s) on which the measurement is to be made.
The CONFigure commands do not make the power measurement after setting the
configuration. Use READ?, or alternatively use INITiate followed by a FETCh? to
make the measurement.
The CONFigure command also applies the following defaults to the channel(s)
which are in the specified window (the channel(s) in the window are specified in
the <source list> parameter):
Defaul t Settings
INITiate:CONTinuous OFF
TRIGger:SOURce IMMediate
TRIGger:DELay:AUTO ON
SENSE:AVERage:COUNt:AUTO ON
SENSE:AVERage:STATe ON
N1911A/1912A P-Series Power Meters Programming Guide
MEASurement Commands
Description
Sets the power meter to make one trigger cycle when INITiate is
sent.
When TRIG:SOUR is set to BUS or HOLD, sets the power meter
to make the measurement immediately a trigger is received.
Enables automatic delay before making the measurement.
Enables automatic filter length selection.
Enables averaging.