Command
[SENS[1]]|SENS2:SWE[1]|2|3|4
:TIME
Trigger setup
TRIG[:SEQ[1]|2]:SOUR
INIT:CONT
TRIG[:SEQ]:LEV:AUTO
TRIG[:SEQ]:LEV
TRIG[:SEQ]:SLOP
TRIG[:SEQ]:DEL
TRIG[:SEQ]:HOLD
[a]
Range
[SENS[1]]|SENS2:POW:AC:RANG:AUTO
[SENS[1]]|SENS2:POW:AC:RANG
Step detection
[SENSe[1]]|SENS2:AVER:SDET
[a] The Range setting in
Table 12-8
N1911A/1912A P-Series Power Meters Programming Guide
Setting
Gate 1: 10 ms
Gates 2 - 4: 0
INT1
ON
ON
AUTO
POS
0 s
MIN
OFF
UPPER
0
is only applicable for E-Series power sensor and N8480 Series power sensor (excluding Option CFT).
SYSTem Subsystem
Comments
Length of time gated period for time gated
measurements.
Trigger source set up and acquisition
mode continuous triggering
Enable automatic setting of the trigger
level
Automatic Power level
Trigger event recognized on the rising
edge of a signal
Delay between recognition of trigger event
and start of a measurement
Trigger holdoff
Auto range off
Range set to upper
Step detection disabled
12
543