Measure[1]|2|3|4[:Scalar][:Power:ac]:Ratio? [<Expected_Value>[,<Resolution>[,<Source List>]]] - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

2
MEASurement Commands
MEASure[1]|2|3|4[:SCALar][:POWer:AC]:RATio?
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the specified window's measurement function to ratio mode
with relative mode off, aborts, configures then initiates both Channel A and B,
calculates the ratio measurement result and places the result on the bus. The
result is a ratio based measurement and is expressed in the units defined by
UNIT[1]|2|3|4:POWer:RATio.
Syntax
MEAS
1
2
3
4
Parameters
Refer to
parameters in this command.
Item
expected_value
(for the expected power
level)
resolution
170
:POW
:SCAL
expected_value
Space
DEF
"Optional Parameters"
Description/Defaul t
The power meter ignores the numeric value entered in this
parameter. Any value entered is treated like DEF.
A numeric value for the resolution. If unspecified the current
resolution setting is used.
N1911A/1912A P-Series Power Meters Programming Guide
:AC
:RAT
,
resolution
DEF
on page 112 for additional details on the
?
,
source list
Range of Values
sensor dependent
[a]
DEF
[b]
1 to 4
1.0, 0.1, 0.01, 0.001
[a]
DEF

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1912a

Table of Contents