Sense[1]]|Sense2:Sweep[1]|2|3|4:Auto:ref1|Ref2 <Numeric_Value - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

10
SENSe Subsystem
Example
SENS:SWE2:AUTO ON
SENS2:SWE3:AUTO OFF
SENS2:SWE4:AUTO ONCE
Reset Condition
On reset, Perpetual Gating will be disabled.
Query
[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:AUTO?
The query returns the current setting of perpetual gating(0 or 1).
– 1 is returned if the perpetual gating is turned on
– 0 is returned if the perpetual gating is turned off
Query Example
SENS2:SWE:AUTO?
Error Messages
– If no sensor or wrong sensor is connected to the channel, error –241
– If channel is not in trigger acquisition mode, error –221 "Settings conflict"
– If Auto Gate fails, error –221 "Settings conflict; Auto Once failed" occurs.
436
"Hardware missing" occurs.
occurs.
N1911A/1912A P-Series Power Meters Programming Guide
This command turns on Channel A Gate 2
Perpetual Gating.
This command turns off Channel B Gate 3
Perpetual Gating.
This command triggers Auto Gating for
Channel B Gate 4.
The query returns the current setting of
Perpetual Gating for Gate 1 Channel B.