Service:option <Character_Data - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

SERVice:OPTion <character_data>
This command loads the power meter memory with the options fitted. The query
form of the command can be used to determine which options are fitted to the
unit.
Syntax
SERV
Parameters
Item
character_data
Example
SERV:OPT "003"
Query
SERVice:OPTion?
The query returns the current option string. For example, if the string "003" is
returned, the power meter is fitted with a sensor input and power reference on the
back panel.
N1911A/1912A P-Series Power Meters Programming Guide
Space
:OPT
Description/Defaul t
Details the option number in a comma separated list.
A maximum of 30 characters can be used.
character_data
?
This command loads the power meter
memory with 003 indicating that the unit is a
rear panel option.
SERVice Subsystem
Range of Values
A to Z (uppercase)
a to z (lowercase)
0 - 9
_ (underscore)
15
701

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1912a

Table of Contents