Table 5-4 Measurement Units; Display[:Window[1]|2]:Meter:upper <Numeric_Value - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

DISPlay[:WINDow[1]|2]:METer:UPPer <numeric_value>
This command sets the analog meter upper scale limit.
This command has the same purpose as DISPlay[:WINDow[1]|2]:ANALog:UPPer
NOTE
<numeric_value>.
The units used are dependent on the current setting of UNIT:POWer and
CALCulate:RELative:STATe as shown in
Table 5-4
Measurement Units
Measurement Mode
Single Channel
Ratio
Difference
N1911A/1912A P-Series Power Meters Programming Guide
Measurement Type
Avg, Pk
Pk–Avg
Avg, Pk, Pk–Avg
Avg, Pk
Pk–Avg
Table
CALC:REL:STAT OFF
Linear
Log
Watt
dBm
%
dB
%
dB
Watt
dBm
%
dB
DISPlay Subsystem
5-4.
CALC:REL:STAT ON
Linear
%
%
%
%
%
5
Log
dB
dB
dB
dB
dB
249

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1912a

Table of Contents