Service:sensor[1]|2:Radc - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

SERVice:SENSor[1]|2:RADC?

This query returns a new raw uncorrected measurement in volts, as a 32 bit signed
integer.
For E9320 Series sensors:
NOTE
Before running this query, the voltage to corrected power conversion must be
disabled using the SERVice:SENSor[1]|2:CORRections:STATe command.
Syntax
SERV
Example
SERV:SENS2:RADC?
Error Messages
– If INIT:CONT is set to ON, error –221 "Settings Conflict" occurs.
– If an E9320 Series sensor is connected and
N1911A/1912A P-Series Power Meters Programming Guide
:SENS
SERVice:SENSor[1]|2:CORRections:STATe is set to ON, error –221
"Settings Conflict" occurs.
:RADC
1
2
This query returns a new raw uncorrected
measurement for the sensor connected to
Channel B.
SERVice Subsystem
?
15
719

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1912a

Table of Contents