Configure[1]|2|3|4[:Scalar][:Power:ac]:Ratio: Relative[<Expected_Value>[,<Resolution>[,<Source List>]]] - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

2
MEASurement Commands
CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:RATio:
RELative[<expected_value>[,<resolution>[,<source list>]]]
This command sets the measurement function, range and resolution of the
specified window. It sets the measurement function to ratio with relative mode on.
The relative value used is that set by the CALCulate:RELative:MAGNitude:AUTO
command.
Syntax
CONF
1
2
3
4
Parameters
Refer to
parameters in this command.
Item
expected_value
(for the expected power level)
resolution
source list
130
:POW
:SCAL
expected_value
Space
DEF
"Optional Parameters"
Description/Defaul t
The power meter ignores the numeric value entered in
this parameter. Any value entered is treated like DEF.
A numeric value for the resolution. If unspecified the
current resolution setting is used.
This channel list specifies the channels used to
calculate the ratio.
If unspecified and the current window setup is a ratio
measurement then this ratio setup is used, otherwise it
defaults to Channel A/B (N1912A) or A/A (N1911A).
N1911A/1912A P-Series Power Meters Programming Guide
:RAT
:AC
,
resolution
DEF
on page 112 for additional details on the
:REL
,
source list
Range of Values
sensor dependent
[a]
DEF
[b]
1 to 4
1.0, 0.1, 0.01, 0.001
[a]
DEF
[c]
(@1),(@2)
[c]
(@2),(@1)
(@1),(@1)
[c]
(@2),(@2)