Keysight N1911A Programming Manual page 459

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

Parameters
Item
numeric_value
Example
SENS:TRAC:LIM:UPP 100
Reset Condition
On reset, the value is set to DEF.
Query
SENSe[1]|2:TRACe:LIMit:LOWer [MIN|MAX]
The query returns the current setting of the trace upper scale limit or the value
associated with MIN or MAX. The format of the response is <NR3>.
Query Example
SENS:TRAC:LIM:UPP?
N1911A/1912A P-Series Power Meters Programming Guide
Description/Defaul t
A numeric value for the trace lower scale limit.
: the default is 20 dBm
DEF
: –150 dBm
MIN
: 230 dBm
MAX
Range of Values
–150 to 230 dBm
DEF
MIN
MAX
This command sets the trace upper scale
limit to 100 dBm for Channel A.
This command queries the trace upper scale
limit of Channel A.