TRACe Subsystem
TRACe[1]|2:DEFine:TRANsition:REFerence <numeric_value>,
<numeric_value>
This command defines the reference levels to be used in the calculation of
transition durations and occurrences. This allows transition measurements
between non-standard reference levels and it is a configuration command that
independent of sensors.
Syntax
TRAC
Parameters
Item
Description/Defaul t
numeric_value
Reference levels to be used in calculation of transition durations and occurences
Example
618
:TRAN
1
:DEF
2
TRAC1:DEF:TRAN:REF 1,18
TRAC1:DEF:TRAN:REF DEF,DEF This command sets trace 1 transition
N1911A/1912A P-Series Power Meters Programming Guide
numeric_value
Space
:REF
This command sets trace 1 transition
measurements to look for the 1 % and 81 %
reference levels.
measuremetns to look for the 10 % and 90 %
reference levels.
numeric_value
?
Range of Values
0 to 100