Pstatistic[1]|2:Ccdf:continuous <Boolean - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

PSTatistic[1]|2:CCDF:CONTinuous <boolean>
This command is used to turn on or off the CCDF Continuous Refresh mode for
Channel A or Channel B.
This command is only applicable when P-Series sensor is present and free-run
NOTE
acquisition mode is selected. If P-Series sensor is used by the adjacent channel
in a dual channel setup, the same setting will be applied for free-run acquisition
mode.
Syntax
Example
PST1:CCDF:CONT ON
PST2:CCDF:CONT OFF
Reset Condition
On reset, the CCDF Continuous Refresh mode will be turned on.
N1911A/1912A P-Series Power Meters Programming Guide
:CCDF
1
PST
2
PSTatistic Subsystem
Space
:CONT
This command turns on the CCDF
Continuous Refresh mode for Channel A.
This command turns off the CCDF
Continuous Refresh mode for Channel B and
the Single Refresh mode is on.
0|OFF
1|ON
?
339