Example
BUFF:COUN 100
Query
[SENSe[1]]|SENSe2:BUFFer:COUNt?
This query is used to retrieve the average or peak measurement buffer size.
Query Example
BUFF:COUN?
On Reset
On *RST, the value is set to 1.
Error Messages
– If no sensor or wrong sensor is connected to the channel, error –241,
– If E4410, N8480, E9300, E9320 or N1920 sensor is connected but acquisition
– If frequency sweep step is non-zero, error –221, "Settings conflict. Frequency
– If parameter is set lower than 1, error –222 "Data out of range; value clipped to
– If parameter is set higher than 2048, error –222 "Data out of range; value
N1911A/1912A P-Series Power Meters Programming Guide
"Hardware missing" occurs.
mode is in free run, error –221, "Setting conflict. Invalid acquisition mode"
occurs.
sweep enabled. Buffer count overidden" occurs.
lower limit" occurs.
clipped to upper limit" occurs.
This command sets the average or peak
measurement buffer size to 100 for Channel A.
This query returns the average or peak
measurement buffer size for Channel A.
SENSe Subsystem