TRIGger Subsystem
TRIGger[:SEQuence[1]|2]:DELay:AUTO <boolean>
This command is used to determine whether or not there is a settling-time delay
before a measurement is made.
When this command is set to:
– ON, the power meter inserts a settling-time delay before taking the requested
– OFF, no settling-time delay is inserted and the power meter makes the
– ONCE, a settling-time delay is inserted before taking the requested
TRIGger[:SEQuence[1]|2]:DELay:AUTO is ignored if
TRIGger[1]|2[:IMMediate] is used to trigger the power meter.
Syntax
TRIG
674
measurement and for subsequent measurements. This settling time allows the
internal digital filter to be updated with new values to produce valid, accurate
measurement results. The trigger with delay command allows settling time for
the internal amplifiers and filters. It does not allow time for power sensor
delay.
In cases of large power changes, the delay may not be sufficient for complete
settling. Accurate readings can be assured by taking two successive
measurements for comparison.
measurement immediately a trigger is received.
measurement, for one measurement only.
:SEQ
1
2
N1911A/1912A P-Series Power Meters Programming Guide
:DEL
:AUTO
Space
0|OFF
1|ON
ONCE
?