Table 12-2 Gsm900: Power Meter Presets - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

12
SYSTem Subsystem
GSM900
Table 12-2
GSM900.
The GSM900 set-up provides the following:
– Average power measurement in one GSM timeslot
– Trace display showing "on" timeslot
A GSM900 measurement is started by detecting the rising edge of a GSM RF
burst—for example the burst emitted by a GSM mobile—using the internal RF level
trigger. The trigger level is set to –20 dBm. Time-gating is used to measure the
average power in the useful part of a GSM burst.
Commands not listed are preset according to their DEFault values (for further
information refer to
Table 12-2
GSM900: Power Meter Presets
Command
Frequency
[SENS[1]]|SENS2:FREQ[:CW|:FIX]
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC
Sensor video bandwidth setup
[SENS[1]]|SENS2:BAND|BWID:VID
Gate Setup
[SENS[1]]|SENS2:SWE[1]|2|3|4
:OFF:TIME
[SENS[1]]|SENS2:SWE[1]|2|3|4
:TIME
Trigger setup
TRIG[:SEQ[1]|2]:SOUR
INIT:CONT
534
shows the power meter presets when <character_data> is set to
Table
12-1.
Setting
+900.000 MHz
NORM
E9321A/25A: HIGH
E9322A/26A: MED
E9323A/27A: LOW
N1921/2A: LOW
Gate 1: 20 µs
Gates 2 - 4: 0
Gate 1: 520 µs
Gates 2 - 4: 0
INT1
ON
N1911A/1912A P-Series Power Meters Programming Guide
Comments
Frequency setting
Measurement mode
Sensor video bandwidth
Delay between trigger point and time
gated period.
Length of time gated period for time gated
measurements.
Trigger source set up and acquisition
mode continuous triggering