11
STATus Subsystem

STATus:OPERation:ULFail[:SUMMary]

The operation status upper limit fail summary register set contains information on
the upper limit fail status of the power meter.
The following bits in these registers are used by the power meter:
Bit Number
Decimal Weight
0
-
1
2
2
4
3
8
4
16
5
32
6
64
7-15
-
The appropriate bits are set if a channel upper limit test fails or a window upper
limit test fails.
These bits are cleared if a measurement is made and the test is enabled and
passes.
Syntax
STAT
482
Definition
Not used
Channel A ULFail Status
Channel B ULFail Status (N1912A only)
Upper window ULFail Status
Lower window ULFail Status
Upper window lower measurement LLFail Status
Lower window lower measurement LLFail Status
Not used
:OPER
:ULF
N1911A/1912A P-Series Power Meters Programming Guide
:SUMM