Keysight N1911A Programming Manual page 423

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

Example
FREQ:STEP 10
Query
[SENSe[1]]|SENSe2:FREQuency[:CW|:FIXed]:STEP?
This query is used to retrieve the number of steps in average or peak frequency
sweep.
Query Example
FREQ:STEP?
On Reset
On *RST, the value is set to 0.
Error Messages
– If no sensor or wrong sensor is connected to the channel, error –241,
– If E4410, N8480 (excluding Option CFT), E9300, E9320 or N1920 sensor is
– If parameter set is lower than 0, error –222, "Data out of range; value clipped
– If parameter set is higher than 2048, error –222, "Data out of range; value
N1911A/1912A P-Series Power Meters Programming Guide
"Hardware missing" occurs.
connected but acquisition mode is in free run, error –221, "Setting conflict.
Invalid acquisition mode" occurs.
to lower limit" occurs.
clipped to upper limit" occurs.
This command sets frequency sweep with 10
steps for Channel A.
This query returns the number of steps in
frequency sweep for Channel A.
SENSe Subsystem
10
423