TRIGger Subsystem

INITiate[:IMMediate]:SEQuence[1]|2

This command sets the power meter in the wait for trigger state. When a trigger is
received, the measurement is taken and the result placed in the power meter
memory. If TRIGger:SOURce is set to IMMediate the measurement begins as
soon as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer. Refer
to
This command performs the same function as INITiate[1]|2:[IMMediate].
NOTE
Syntax
Example
INIT:IMM:SEQ1
Error Messages
If the power meter is not in the "idle" state or INITiate:CONTinuous is ON, error
–213, "INIT ignored" occurs.
652
"FETCh[1]|2|3|4 Queries"
INIT
:IMM
N1911A/1912A P-Series Power Meters Programming Guide
on page 132 for further information.
:SEQ
1
2
This command places Channel A in the
wait for trigger state.