[Sense[1]]|Sense2:Correction:cset[1]|Cset2:State <Boolean - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe <boolean>
This command is to enable and disable the use of the currently active sensor
calibration table (CSET1) or frequency dependent offset table (CSET2). When a
table has been selected and enabled, the calibration factors/offsets stored in it
can be used by specifying the required frequency using the
[SENSe[1]]|SENSe2:FREQuency command.
When the CSET1 command is set to ON, the reference calibration factor is taken
from the sensor calibration table and is used during calibration.
Syntax
SENS
1
SENS2
Example
CORR:CSET1:STAT 1
Reset Condition
On reset, the sensor calibration table and frequency dependent offset table are
not affected.
N1911A/1912A P-Series Power Meters Programming Guide
:
CORR
:CSET
:CSET2
1
:STAT
This command enables the use of the
currently active sensor calibration table for
Channel A.
SENSe Subsystem
Space
0|OFF
1|ON
?
10
399

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1912a

Table of Contents