Configure[1]|2|3|4[:Scalar][:Power:ac]:Relative [<Expected_Value>[,<Resolution>[,<Source List>]]] - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

2
MEASurement Commands
CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:RELative
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the measurement function, range and resolution of the
specified window. It sets the measurement function to single channel with relative
mode on. The relative value used is that set by the
CALCulate:RELative:MAGNitude:AUTO command.
Syntax
CONF
1
2
3
4
Parameters
Refer to
parameters in this command.
Item
expected_value
(for the expected power level)
resolution
122
:POW
:SCAL
expected_value
Space
DEF
"Optional Parameters"
Description/Defaul t
A numeric value for the expected power level. The units
of measurement are dBm and W. The default units are
defined by UNIT:POWer.
A numeric value for the resolution. If unspecified the
current resolution setting is used.
N1911A/1912A P-Series Power Meters Programming Guide
:REL
:AC
,
resolution
DEF
on page 112 for additional details on the
,
source list
Range of Values
sensor dependent
[a]
DEF
[b]
1 to 4
1.0, 0.1, 0.01, 0.001