9
PSTatistic Subsystem

PSTatistic[1]|2:CCDF:TRACe:POWer:AVERage?

This command is used to retrieve average power value of Channel A or Channel B
trace.
This command is only applicable when P-Series sensors are present and the
NOTE
following conditions are met:
– Free-run acquisition mode is selected
– NORMal or DOUBle measurement speed setting is chosen
Syntax
Example
PST:CCDF:TRAC:POW:AVER?
Error Messages
– If no power sensor is connected, error –241 "Hardware missing" occurs.
– If sensor/sensors connected are not P-Series sensors, error –241 "Hardware
– If measurement speed setting is FAST, error –221 "Settings conflict" occurs.
– If command executed in other window besides CCDF window, error –221
– If the acquisition mode is in continuous triggering or triggering source is set to
360
1
PST
2
missing" occurs.
"Settings conflict: Requires CCDF window" occurs.
either INT1, INT2, or EXT, error –221 "Settings conflict" occurs.
:CCDF
:TRAC
This command returns the average power
value for Channel A trace.
N1911A/1912A P-Series Power Meters Programming Guide